The different functions of the Nox T3 can be tested in order to verify if the device is fully functional. The following test is to check activity on thorax and/or abdomen channels.
Items needed for test:
- Computer with Noxturnal
- Nox T3 Device
- One AA Battery
- RIP Belt
Test:
Setup a recording on the Nox T3 Device in Noxturnal
Insert AA battery and close the battery compartment
Push the Right Arrow button once, or until you see the respiratory signal screen:
Snap the RIP belts ends to the back panel of the device:
To check the thoracic RIP Belt, pull or stretch the thoracic belt just a little bit. Traces should be seen on the thoracic RIP Belt:
If a trace appears on the display of the device as shown in the figure above the channel is active but if it only shows a flat line as shown in the figure below when pulling the belt the channel is inactive and the device needs to be sent in for repair:
To check the abdomen RIP Belt, pull or stretch the abdomen belt just a little bit. Traces should be seen on the abdomen RIP Belt:
If the trace appears as a flat line when pulling the belt the channel is inactive and the device needs to be sent in for repair
For further information please contact support@noxmedical.com