The performance of Nox Medical systems depends on the batteries. To ensure high-quality recordings every time, particular attention should be paid to battery selection and battery care.

Battery recommendations:

For the Nox T3™: 1 AA battery

  • Alkaline batteries can power the device for up to 16 hours of recording. It is recommended to change them before each recording.
  • Lithium batteries can power the device for over 20 hours of recording. They are the optimal choice for multiple night recordings.
  • NiMh batteries with a capacity of 2500mAh or more last for up to 18 hours of recording. Recharge them before every recording. It is recommended to replace them every 6 to 12 months.

For the Nox A1™: 1 AA battery

  • The Nox A1 does not support Alkaline batteries.
  • Lithium batteries are recommended. They can power the device for over 12 hours of recording. They must be replaced between every recording.
  • Recommended NiMh batteries are Maha Powerex™ 2700mAh and Panasonic Eneloop Pro™ 2500mAh. They can power the device for 8 to 10 hours of recording. Recharge them before every recording. It is recommended to replace them every 3 to 6 months.

For the Nonin™ 3150 oximeter: 2 AAA batteries

  • Alkaline batteries can power the device for up to 24 hours of recording. It is recommended to change them every 2 recordings.
  • NiMh batteries can power the device for up to 20 hours of recording. Recharge them before every recording.

Recommended care for rechargeable NiMh batteries:

General recommendations on the use of NiMh batteries:

  • Batteries should be stored in a cool dry place and used regularly.
  • When not in use, batteries should be left on the charger, even if they are fully charged.
  • Replace the batteries every 3 to 6 months if using a Nox A1 and every 6 to 12 months if using a Nox T3.
  • Replace the charger every 2 years.

The Maha Powerex MH-C801D is the recommended charger for use with Nox systems. It can charge up to 8 batteries at the same time (AA and AAA).

Two charging modes are available:

RAPID: default mode once the charger is powered on. Recharges the batteries in 1 hour.

SOFT (recommended mode): maximises battery life and performance. Recharges the batteries in 2 hours. Initiate the SOFT charging mode as follows:

  1. Remove all the batteries from the charger.
  2. Insert one battery into the leftmost slot.
  3. Within five seconds of inserting the battery, press and hold the “SOFT” button until “SOFT CHARGING” is displayed on the screen.
  4. Insert the remaining batteries.

It is recommended to condition batteries every 2 weeks. The conditioning cycle requires up to 14 hours to complete. Initiate the conditioning cycle as follows:

  1. Remove all the batteries from the charger.
  2. Insert one battery into the leftmost slot.
  3. Within five seconds of inserting the battery, press and hold the  button until “CONDITION” is displayed on the screen.
  4. Insert the remaining batteries.

Ref Numbers / Part numbers (PN)*:

*All products listed with PN number (format: 70XXXXX) are reference to the ResMed catalogue. The REF numbers (format: 56 XX XX) refer to products available in the Nox Medical Product catalog. Contact your distributor for more information.

LBL-0066 REV3

 

*Please note that not all features or all products are available in all markets. Our products are medical devices and subject to registration/approval in each market area. For more information on product availability please contact support@noxmedical.com